유방촬영 시 타겟/필터별 평균유선선량 환산인자 도출 (Derivation of the Mean Glandular Dose Conversion Factors for each Target /Filter in Mammography)

10 페이지
기타파일
최초등록일 2025.03.17 최종저작일 2022.03
10P 미리보기
유방촬영 시 타겟/필터별 평균유선선량 환산인자 도출
  • 미리보기

    서지정보

    · 발행기관 : (사)한국방사선산업학회
    · 수록지 정보 : 방사선산업학회지 / 16권 / 1호 / 13 ~ 22페이지
    · 저자명 : 허재빈, 남형우, 곽민우, 김민성, 김광표

    초록

    As the incidence of breast cancer in Korea increases, mammography usage is alsoincreasing. Mean glandular dose (MGD) is used to manage radiation dose in mammography. The MGDis calculated using the MGD conversion factor and measurable radiation quantities. The previouslydeveloped MGD conversion factors require half value layer (HVL) that is difficult to measure in medicalinstitutions when used. Therefore, the object of this study is to derive the MGD conversion factor usingexamination parameter that is easy to obtain, such as tube voltage. In this study, the X-ray spectra of themammography device for each target /filter were calculated to derive the MGD conversion factors. Thebreast computational phantom and the exposure situation in mammography were simulated. The MGDconversion factors were derived using the MCNP6 code, and compared with the previously developedMGD conversion factors to verify the derived MGD conversion factors. In the X-ray spectra for eachtarget/filter derived in this study, the mean energy was high in the order of W/Rh, Mo/Rh, and Mo/Mo atthe same tube voltage. As a result of comparing under the same parameters, the MGD conversion factorswere the highest at W/Rh, followed by Mo/Rh and Mo/Mo. When Mo/Rh and Mo/Mo were compared,the MGD conversion factors of Mo/Rh were higher in the all tube voltage. For all targets /filters, theMGD conversion factors increased as the tube voltage increased at the same glandularity, and theMGD conversion factors decreased as the compressed breast thickness increased. Compared to theMGD conversion factors of Dance and Wu, the differences were 8.9% and 6.8%, respectively. The MGDconversion factors developed in this study will contribute to the reduction of radiation doses resultingfrom mammography through use in managing radiation doses resulting from mammography at medicalinstitutions that are difficult to measure HVL.

    영어초록

    As the incidence of breast cancer in Korea increases, mammography usage is alsoincreasing. Mean glandular dose (MGD) is used to manage radiation dose in mammography. The MGDis calculated using the MGD conversion factor and measurable radiation quantities. The previouslydeveloped MGD conversion factors require half value layer (HVL) that is difficult to measure in medicalinstitutions when used. Therefore, the object of this study is to derive the MGD conversion factor usingexamination parameter that is easy to obtain, such as tube voltage. In this study, the X-ray spectra of themammography device for each target /filter were calculated to derive the MGD conversion factors. Thebreast computational phantom and the exposure situation in mammography were simulated. The MGDconversion factors were derived using the MCNP6 code, and compared with the previously developedMGD conversion factors to verify the derived MGD conversion factors. In the X-ray spectra for eachtarget/filter derived in this study, the mean energy was high in the order of W/Rh, Mo/Rh, and Mo/Mo atthe same tube voltage. As a result of comparing under the same parameters, the MGD conversion factorswere the highest at W/Rh, followed by Mo/Rh and Mo/Mo. When Mo/Rh and Mo/Mo were compared,the MGD conversion factors of Mo/Rh were higher in the all tube voltage. For all targets /filters, theMGD conversion factors increased as the tube voltage increased at the same glandularity, and theMGD conversion factors decreased as the compressed breast thickness increased. Compared to theMGD conversion factors of Dance and Wu, the differences were 8.9% and 6.8%, respectively. The MGDconversion factors developed in this study will contribute to the reduction of radiation doses resultingfrom mammography through use in managing radiation doses resulting from mammography at medicalinstitutions that are difficult to measure HVL.

    참고자료

    · 없음
  • 자주묻는질문의 답변을 확인해 주세요

    해피캠퍼스 FAQ 더보기

    꼭 알아주세요

    • 자료의 정보 및 내용의 진실성에 대하여 해피캠퍼스는 보증하지 않으며, 해당 정보 및 게시물 저작권과 기타 법적 책임은 자료 등록자에게 있습니다.
      자료 및 게시물 내용의 불법적 이용, 무단 전재∙배포는 금지되어 있습니다.
      저작권침해, 명예훼손 등 분쟁 요소 발견 시 고객센터의 저작권침해 신고센터를 이용해 주시기 바랍니다.
    • 해피캠퍼스는 구매자와 판매자 모두가 만족하는 서비스가 되도록 노력하고 있으며, 아래의 4가지 자료환불 조건을 꼭 확인해주시기 바랍니다.
      파일오류 중복자료 저작권 없음 설명과 실제 내용 불일치
      파일의 다운로드가 제대로 되지 않거나 파일형식에 맞는 프로그램으로 정상 작동하지 않는 경우 다른 자료와 70% 이상 내용이 일치하는 경우 (중복임을 확인할 수 있는 근거 필요함) 인터넷의 다른 사이트, 연구기관, 학교, 서적 등의 자료를 도용한 경우 자료의 설명과 실제 자료의 내용이 일치하지 않는 경우
문서 초안을 생성해주는 EasyAI
안녕하세요. 해피캠퍼스의 방대한 자료 중에서 선별하여 당신만의 초안을 만들어주는 EasyAI 입니다.
저는 아래와 같이 작업을 도와드립니다.
- 주제만 입력하면 목차부터 본문내용까지 자동 생성해 드립니다.
- 장문의 콘텐츠를 쉽고 빠르게 작성해 드립니다.
- 스토어에서 무료 캐시를 계정별로 1회 발급 받을 수 있습니다. 지금 바로 체험해 보세요!
이런 주제들을 입력해 보세요.
- 유아에게 적합한 문학작품의 기준과 특성
- 한국인의 가치관 중에서 정신적 가치관을 이루는 것들을 문화적 문법으로 정리하고, 현대한국사회에서 일어나는 사건과 사고를 비교하여 자신의 의견으로 기술하세요
- 작별인사 독후감
해캠 AI 챗봇과 대화하기
챗봇으로 간편하게 상담해보세요.
2025년 07월 20일 일요일
AI 챗봇
안녕하세요. 해피캠퍼스 AI 챗봇입니다. 무엇이 궁금하신가요?
9:45 오후