Ti-6Al-4V 합금 미니 임플란트에 마찰고정한 경사 지대주의 피로특성 (Fatigue Characteristics of Ti-6Al-4V Alloy Mini-Implant with Friction Locked Abutment)

10 페이지
기타파일
최초등록일 2025.06.22 최종저작일 2008.12
10P 미리보기
Ti-6Al-4V 합금 미니 임플란트에 마찰고정한 경사 지대주의 피로특성
  • 미리보기

    서지정보

    · 발행기관 : 대한치과재료학회
    · 수록지 정보 : 대한치과재료학회지 / 35권 / 4호 / 359 ~ 368페이지
    · 저자명 : 황갑운, 이만형, 송경우, 이민호, 배태성

    초록

    Titanium miniscrew implant systems are being used as a temporary implant system, because
    they are easy to place and to remove. The use of titanium miniscrew implant systems as a semi-permanent implant has not been reported in clinical case reports. To our knowledge, there
    are no published reports evaluating the stability of mini implant systems. Information about the
    shape of mini implants used in relation to the stability vary depending on bone quality.
    This study was carried to find out the possibility of semi-permanent use of the 15°-angled
    abutment friction-locked on Ti-6Al-4V alloy implant fixtures with diameter of 2.5, 3.0 and 3.5
    mm. At first, we measured the pull-out strengths of the abutment friction-locked by 4 strokes of
    mallet with the weight of 130 g. Secondly, we measured pull-out strengths of the abutment after
    the cyclic compression loading of the amplitude range of 2 to 100 N during 1×106 cycles. Cyclic
    loading was performed based on ISO/FDIS 14801:2003(E). Test specimen was clamped to be 10
    degree with the loading axis of the testing machine. The cyclic loading was applied at load cycles of 10 Hz using dynamic fatigue testing machine. Fatigue test was set the amplitude range to be within 2 % to 80 % of the static strength. Fatigue limit that guarantee a life of 5×106 cycles.
    The results obtained were summarized as follows;
    1. The retention force was increased by friction locking between implant fixture and abutment
    during the cyclic loading with compression and bending.
    2. Fatigue limits measured in this study were 261.6 N, 303.1 N, and 552.2 N for 2.5 ㎜, 3.0 ㎜,
    and 3.5 ㎜ in diameter of implant, respectively. It means that the fatigue limits tend to increase with the cross sectioned area of implant.
    3. Failures of implants were not observed at the connected area between implant fixture and
    abutment. However, bending and/or fracture was happened at the screw area of implant
    fixture. Judging from the above results, it will be good stability for clinical application, when the
    15°-angled abutment friction-locked on mini implant fixture of diameter of 2.5 ㎜ are properly used for anterior prosthetic restoration.

    영어초록

    Titanium miniscrew implant systems are being used as a temporary implant system, because
    they are easy to place and to remove. The use of titanium miniscrew implant systems as a semi-permanent implant has not been reported in clinical case reports. To our knowledge, there
    are no published reports evaluating the stability of mini implant systems. Information about the
    shape of mini implants used in relation to the stability vary depending on bone quality.
    This study was carried to find out the possibility of semi-permanent use of the 15°-angled
    abutment friction-locked on Ti-6Al-4V alloy implant fixtures with diameter of 2.5, 3.0 and 3.5
    mm. At first, we measured the pull-out strengths of the abutment friction-locked by 4 strokes of
    mallet with the weight of 130 g. Secondly, we measured pull-out strengths of the abutment after
    the cyclic compression loading of the amplitude range of 2 to 100 N during 1×106 cycles. Cyclic
    loading was performed based on ISO/FDIS 14801:2003(E). Test specimen was clamped to be 10
    degree with the loading axis of the testing machine. The cyclic loading was applied at load cycles of 10 Hz using dynamic fatigue testing machine. Fatigue test was set the amplitude range to be within 2 % to 80 % of the static strength. Fatigue limit that guarantee a life of 5×106 cycles.
    The results obtained were summarized as follows;
    1. The retention force was increased by friction locking between implant fixture and abutment
    during the cyclic loading with compression and bending.
    2. Fatigue limits measured in this study were 261.6 N, 303.1 N, and 552.2 N for 2.5 ㎜, 3.0 ㎜,
    and 3.5 ㎜ in diameter of implant, respectively. It means that the fatigue limits tend to increase with the cross sectioned area of implant.
    3. Failures of implants were not observed at the connected area between implant fixture and
    abutment. However, bending and/or fracture was happened at the screw area of implant
    fixture. Judging from the above results, it will be good stability for clinical application, when the
    15°-angled abutment friction-locked on mini implant fixture of diameter of 2.5 ㎜ are properly used for anterior prosthetic restoration.

    참고자료

    · 없음
  • 자주묻는질문의 답변을 확인해 주세요

    해피캠퍼스 FAQ 더보기

    꼭 알아주세요

    • 자료의 정보 및 내용의 진실성에 대하여 해피캠퍼스는 보증하지 않으며, 해당 정보 및 게시물 저작권과 기타 법적 책임은 자료 등록자에게 있습니다.
      자료 및 게시물 내용의 불법적 이용, 무단 전재∙배포는 금지되어 있습니다.
      저작권침해, 명예훼손 등 분쟁 요소 발견 시 고객센터의 저작권침해 신고센터를 이용해 주시기 바랍니다.
    • 해피캠퍼스는 구매자와 판매자 모두가 만족하는 서비스가 되도록 노력하고 있으며, 아래의 4가지 자료환불 조건을 꼭 확인해주시기 바랍니다.
      파일오류 중복자료 저작권 없음 설명과 실제 내용 불일치
      파일의 다운로드가 제대로 되지 않거나 파일형식에 맞는 프로그램으로 정상 작동하지 않는 경우 다른 자료와 70% 이상 내용이 일치하는 경우 (중복임을 확인할 수 있는 근거 필요함) 인터넷의 다른 사이트, 연구기관, 학교, 서적 등의 자료를 도용한 경우 자료의 설명과 실제 자료의 내용이 일치하지 않는 경우

“대한치과재료학회지”의 다른 논문도 확인해 보세요!

문서 초안을 생성해주는 EasyAI
안녕하세요 해피캠퍼스의 20년의 운영 노하우를 이용하여 당신만의 초안을 만들어주는 EasyAI 입니다.
저는 아래와 같이 작업을 도와드립니다.
- 주제만 입력하면 AI가 방대한 정보를 재가공하여, 최적의 목차와 내용을 자동으로 만들어 드립니다.
- 장문의 콘텐츠를 쉽고 빠르게 작성해 드립니다.
- 스토어에서 무료 이용권를 계정별로 1회 발급 받을 수 있습니다. 지금 바로 체험해 보세요!
이런 주제들을 입력해 보세요.
- 유아에게 적합한 문학작품의 기준과 특성
- 한국인의 가치관 중에서 정신적 가치관을 이루는 것들을 문화적 문법으로 정리하고, 현대한국사회에서 일어나는 사건과 사고를 비교하여 자신의 의견으로 기술하세요
- 작별인사 독후감
해캠 AI 챗봇과 대화하기
챗봇으로 간편하게 상담해보세요.
2026년 02월 04일 수요일
AI 챗봇
안녕하세요. 해피캠퍼스 AI 챗봇입니다. 무엇이 궁금하신가요?
3:43 오후