비침습적 산소 및 이산화탄소 분압의 감시 - 경피적 산소 및 이산화탄소 분압과 호기말 이산화탄소 분압 측정의 정확도 - (Non-invasive Monitoring of Oxygen and Carbon Dioxide Tension - Accuracy of Transcutaneous O₂ and CO₂ and Endtidal CO₂ -)

7 페이지
기타파일
최초등록일 2025.04.28 최종저작일 2009.10
7P 미리보기
비침습적 산소 및 이산화탄소 분압의 감시 - 경피적 산소 및 이산화탄소 분압과 호기말 이산화탄소 분압 측정의 정확도 -
  • 미리보기

    서지정보

    · 발행기관 : 대한응급의학회
    · 수록지 정보 : 대한응급의학회지 / 20권 / 5호 / 481 ~ 487페이지
    · 저자명 : 정동민, 문성우, 최성혁, 홍윤식, 이성우, 임성익, 윤영훈

    초록

    Purpose: To determine the concordance of transcutaneous
    CO₂ (PtcCO₂) versus arterial CO₂ (PaCO₂), end-tidal CO₂
    (PetCO₂) versus PaCO₂, and transcutaneous O₂ (PtcO₂)
    versus arterial O₂ (PaCO₂) among healthy adult volunteers,
    and to determine the normal values of the PtcCO₂/PtcO₂
    and PtcO₂/PaO₂ that will be used as early signs of shock or
    as prognostic factors for critically ill patients.
    Methods: We measured the PtcO₂, PtcCO₂, PetCO₂, PaO₂,
    and PaCO₂ from 11 healthy volunteers while breathing
    room air or O₂ at a flow rate of 6 L/min via nasal cannula.
    The PtcO₂ and PtcCO₂ were measured using a
    Radiometer’'s transcutaneous sensor that interfaced with
    the Solar 8000 patient monitor system. The PetCO₂ was
    measured using a side stream capnometer that sampled air
    from a nasal catheter. The PaO₂ and PaCO₂ were measured
    from arterial blood samples. The concordances of the
    PtcCO₂ versus the PaCO₂, the PtcO₂ versus the PaO₂, and
    the PetCO₂ versus the PaCO₂ were analyzed using a
    Bland-Altman plot. We defined the normal values of the
    P(a-tc)CO₂, PtcO₂/PaO₂, and PtcCO₂/PtcO₂.
    Results: Twenty-two pairs of the PtcCO₂ versus PaCO₂,
    PtcO₂ versus PaO₂, and PetCO₂ versus PaCO₂ were
    obtained. The mean (±SD) values of the P(a-tc)CO₂, P(atc)
    O₂, and P(a-et)CO₂ were 0(±2.2) mmHg, 35.4(±24.1)
    mmHg, and 1.4(±1.3) mmHg, respectively (p=0.947,
    p<0.001, and p<0.001 by paired t-test, respectively). The
    P(a-tc)CO₂ and P(a-et)CO₂ showed a high concordance of
    95.5% within a range of ±4 mmHg. The median (25~75%)
    values of the PtcCO₂/PtcO₂ and PtcO₂/PaO₂ at room air
    were 54.8%(46.8%~6₂.7%), respectively.
    Conclusion: The PtcCO₂ and PetCO₂ had a reliable concordance
    with the PaCO₂. However, the PtcO₂ was discordant
    with the PaO₂ and this discordance was increased when
    inspiring O₂. Therefore, the absolute values of the PtcO₂ cannot
    be used as a surrogate measurement of the PaO₂.
    However, because the O₂ supply did not increase the
    PtcCO₂, but rather the PtcO₂, we can use the trend in the
    change in the PtcCO₂/PtcO₂ or PtcO₂/PaO₂ in shock patients

    영어초록

    Purpose: To determine the concordance of transcutaneous
    CO₂ (PtcCO₂) versus arterial CO₂ (PaCO₂), end-tidal CO₂
    (PetCO₂) versus PaCO₂, and transcutaneous O₂ (PtcO₂)
    versus arterial O₂ (PaCO₂) among healthy adult volunteers,
    and to determine the normal values of the PtcCO₂/PtcO₂
    and PtcO₂/PaO₂ that will be used as early signs of shock or
    as prognostic factors for critically ill patients.
    Methods: We measured the PtcO₂, PtcCO₂, PetCO₂, PaO₂,
    and PaCO₂ from 11 healthy volunteers while breathing
    room air or O₂ at a flow rate of 6 L/min via nasal cannula.
    The PtcO₂ and PtcCO₂ were measured using a
    Radiometer’'s transcutaneous sensor that interfaced with
    the Solar 8000 patient monitor system. The PetCO₂ was
    measured using a side stream capnometer that sampled air
    from a nasal catheter. The PaO₂ and PaCO₂ were measured
    from arterial blood samples. The concordances of the
    PtcCO₂ versus the PaCO₂, the PtcO₂ versus the PaO₂, and
    the PetCO₂ versus the PaCO₂ were analyzed using a
    Bland-Altman plot. We defined the normal values of the
    P(a-tc)CO₂, PtcO₂/PaO₂, and PtcCO₂/PtcO₂.
    Results: Twenty-two pairs of the PtcCO₂ versus PaCO₂,
    PtcO₂ versus PaO₂, and PetCO₂ versus PaCO₂ were
    obtained. The mean (±SD) values of the P(a-tc)CO₂, P(atc)
    O₂, and P(a-et)CO₂ were 0(±2.2) mmHg, 35.4(±24.1)
    mmHg, and 1.4(±1.3) mmHg, respectively (p=0.947,
    p<0.001, and p<0.001 by paired t-test, respectively). The
    P(a-tc)CO₂ and P(a-et)CO₂ showed a high concordance of
    95.5% within a range of ±4 mmHg. The median (25~75%)
    values of the PtcCO₂/PtcO₂ and PtcO₂/PaO₂ at room air
    were 54.8%(46.8%~6₂.7%), respectively.
    Conclusion: The PtcCO₂ and PetCO₂ had a reliable concordance
    with the PaCO₂. However, the PtcO₂ was discordant
    with the PaO₂ and this discordance was increased when
    inspiring O₂. Therefore, the absolute values of the PtcO₂ cannot
    be used as a surrogate measurement of the PaO₂.
    However, because the O₂ supply did not increase the
    PtcCO₂, but rather the PtcO₂, we can use the trend in the
    change in the PtcCO₂/PtcO₂ or PtcO₂/PaO₂ in shock patients

    참고자료

    · 없음
  • 자주묻는질문의 답변을 확인해 주세요

    해피캠퍼스 FAQ 더보기

    꼭 알아주세요

    • 자료의 정보 및 내용의 진실성에 대하여 해피캠퍼스는 보증하지 않으며, 해당 정보 및 게시물 저작권과 기타 법적 책임은 자료 등록자에게 있습니다.
      자료 및 게시물 내용의 불법적 이용, 무단 전재∙배포는 금지되어 있습니다.
      저작권침해, 명예훼손 등 분쟁 요소 발견 시 고객센터의 저작권침해 신고센터를 이용해 주시기 바랍니다.
    • 해피캠퍼스는 구매자와 판매자 모두가 만족하는 서비스가 되도록 노력하고 있으며, 아래의 4가지 자료환불 조건을 꼭 확인해주시기 바랍니다.
      파일오류 중복자료 저작권 없음 설명과 실제 내용 불일치
      파일의 다운로드가 제대로 되지 않거나 파일형식에 맞는 프로그램으로 정상 작동하지 않는 경우 다른 자료와 70% 이상 내용이 일치하는 경우 (중복임을 확인할 수 있는 근거 필요함) 인터넷의 다른 사이트, 연구기관, 학교, 서적 등의 자료를 도용한 경우 자료의 설명과 실제 자료의 내용이 일치하지 않는 경우

“대한응급의학회지”의 다른 논문도 확인해 보세요!

문서 초안을 생성해주는 EasyAI
안녕하세요. 해피캠퍼스의 방대한 자료 중에서 선별하여 당신만의 초안을 만들어주는 EasyAI 입니다.
저는 아래와 같이 작업을 도와드립니다.
- 주제만 입력하면 목차부터 본문내용까지 자동 생성해 드립니다.
- 장문의 콘텐츠를 쉽고 빠르게 작성해 드립니다.
- 스토어에서 무료 캐시를 계정별로 1회 발급 받을 수 있습니다. 지금 바로 체험해 보세요!
이런 주제들을 입력해 보세요.
- 유아에게 적합한 문학작품의 기준과 특성
- 한국인의 가치관 중에서 정신적 가치관을 이루는 것들을 문화적 문법으로 정리하고, 현대한국사회에서 일어나는 사건과 사고를 비교하여 자신의 의견으로 기술하세요
- 작별인사 독후감
해캠 AI 챗봇과 대화하기
챗봇으로 간편하게 상담해보세요.
2025년 08월 03일 일요일
AI 챗봇
안녕하세요. 해피캠퍼스 AI 챗봇입니다. 무엇이 궁금하신가요?
10:00 오전