초음파 가열 시 In Vitro 및 In Vivo에서 Microwave Radio-Thermometer와 탐침온도계의 일치도
* 본 문서는 배포용으로 복사 및 편집이 불가합니다.
서지정보
ㆍ발행기관 : 한국전문물리치료학회
ㆍ수록지정보 : 한국전문물리치료학회지 / 10권 / 1호
ㆍ저자명 : 이수영, 조상현, 이충휘, 김종만
ㆍ저자명 : 이수영, 조상현, 이충휘, 김종만
영어 초록
Therapeutic ultrasound is commonly applied for deep heating in physical therapy setting. However, it is difficult to determine the exact application dosage and to confirm the immediate heating effect. Microwave Radio-Thermometer (MRT) can measure the temperature by the electromagnetic energy in the microwave region of the object that emits above absolute zero temperature. MRT was used for early diagnosis of breast cancer since it was not harmful, non-invasive, and non-ionizing to the human body. The purposes of this study were to investigate how accurately 1.1 GHz RTM (RES Ltd. Russia) measures the change of average temperature in the tissue, and to determine the depth of temperature change measurement. Therapeutic ultrasound was applied (continuous wave for 5 minutes, 1 MHz, intensity of 1.5 [in vitro] and 1.0 [in vivo]) in four different conditions: (1) 30 cases of in vitro specimen of pork, (2) 30 cases of in vitro specimen of pork ankle joint, (3) 10 cases of in vivo canine thigh, and (4) 30 cases of in vivo human body. Intraclass Correlation Coeffients (ICC[3,1]) between average needle probe thermometer below surface and MRT temperature was revealed as followed: (1) Before ultrasound application ICCs ranges above .8 in specimen of pork (15 mm underneath the skin) and above .82 in specimen of pork ankle joint (10~30 mm underneath the skin). (2) After ultrasound application ICCs ranges above .7 in both specimens of pork and pork ankle joint. (3) Before ultrasound application ICCs ranges above .8 in canine thigh (20 mm underneath the skin). (4) After ultrasound application ICCs ranges above .82 in canine thigh. The temperature of the human body increased significantly with the mean of in muscle tissue and with the mean of in joint (p<.00). It was revealed that the average depth of temperature measurement of the tissue by MRT was in between 10 and 35 mm, and determined that the proper temperature measurement band was .참고 자료
없음"한국전문물리치료학회지"의 다른 논문
- 원주시 지역사회중심 재활사업 평가20페이지
- 동결견 환자의 초음파치료 시 신장여부에 따른 관절가동범위 회복과 통증감소 효과 비교9페이지
- 12주 근력강화훈련 동안 표면근전도 신호의 특성20페이지
- 미끄럼 바닥에서 안정성 유지를 위한 균형 전략과 평가방법12페이지
- 정상인에서 보행속도가 발관절의 관절각과 발바닥 최대 압력 분포에 미치는 영향19페이지
- 복횡근 강화운동이 체간 신전-굴곡 시 척추 분절 운동에 미치는 영향14페이지
- 최대 등척성 수축시 표면근전도에서 근 수축 개시점 결정을 위한 기법들의 신뢰도12페이지
- 경피신경전기자극 후 상부 승모근 활성도와 EMG gap의 변화14페이지
- 체중지지 트레드밀훈련이 편마비 환자의 보행과 서기균형에 미치는 영향7페이지
- 낙상 경험 유무에 따른 노인의 기능적 보행성취도 점수(FAP score) 비교13페이지